Research on the calibration method of the avalanche gap of MTPC anode plate
Time projection chamber,Multi-purpose Time Projection Chamber (MTPC),gap calibration,Micromegas,photon feedback
Abstract Accepted
Chu Tianzhi / Chinese Academy of Sciences;Shenzhen University;Institute of high energy physics;China Spallation Neutron Source
Chen Hongkun / Sun Yat-sen University
Han Yi / Institute of High Energy Physics; Chinese Acadamy of Sciense
Haizheng Chen / Institute of High Energy Physics; Chinese Academy of Sciences (CAS)
This report presents an investigation on the calibration of anode plate of the Multi-purpose Time Projection Chamber (MTPC). MTPC mainly acquires the energy of the emitted particles from the responsed pad. The energy information on the pad mainly depends on the gain uniformity. Therefore, it is necessary to calibrate the electron avalanche gain and the energy resolution of MTPC. The Micromegas detector is applied as the anode plate of MTPC, and the avalanche gap between the mesh and the resistive layer directly determine the gain of the anode plate. This report delineates the approach for calibrating the avalanche gap of the anode plate via X-ray source and alpha source. By fitting the curve of elctron avalanche gain with photon feedback coefficient, we obtain the avalanche gap of the anode plate.